OXYGEN DEFICIENCY MONITOR PAGE 13
11. Specifications
SENSOR OXYMAX™ oxygen concentration
transducer.
DISPLAY 3 digit LCD, %0
2
in air from 0% to 25.0%
with low battery indicator.
ACCURACY +
3% of full scale from 10 to 50
o
C
+
4% of full scale from 0 to 50
o
C
RESPONSE 10 seconds to 90% indication
RESOLUTION 0.1% Oxygen
SENSOR LIFE 21 months in ambient air
PRESSURE SENSITIVITY 0.33% 0
2
change for 300 foot elevation
change or 75 mm Hg barometric
change.
TEMPERATURE Operating Range: 0 to 50
o
C